2553 High Voltage Potentiostat-Galvanostat
Potentiostat with large dynamic range
Model 2553 allows for experiments where a large compliance potential is required, along with stability of the measurement and long-time reliability. The new version offers positive feedback control, fast transient response, automatic or manual current ranging and IR compensation.
Detection | |
AD | Amperometric Detection |
PD | Potentiometric Detection |
DSA | Double Step Amperometry |
DSV | Double Step Potentiometry |
PAD | Pulsed Amperometric Detection |
Voltammetric | |
LSV | Linear Scan Voltammetry |
CYV | Cyclic Voltammetry |
GLV | Galvanostatic Linear Voltammetry |
GCV | Galvanostatic Cyclic Voltammetry |
SWV | Square Wave Voltammetry |
NPV | Normal Pulse Voltammetry |
ACV | Alternating Current Voltammetry |
DPV | Differential Pulse Voltammetry |
DNV | Differential Normal Pulse Voltammetry |
DAV | Differential Alternate Pulse Voltammetry |
Stripping | |
LSS | Linear Scan Stripping |
ACS | Alternate Current Stripping |
SWS | Square Wave Stripping |
DAS | Differential Stripping |
DPS | Differential Pulse Stripping |
DNS | Differential Normal Pulse Stripping |
PSA | Potentiometric Stripping Analysis |
CCSA | Constant Current Stripping Analysis |
Specifications are subject to change without previous notice
Counter Electrode | |
Voltage Output | ± 45V max |
Current Output | ± 1,2A max |
Slew Rate | 0,01mV/s to 10V/s |
Protection | Thermal, overload and short-circuit |
Working Electrode | |
Current Measure | From 10nA to 1A Full Scale in 9 ranges |
Current Resolution | From 10pA at 10nA Full Scale to 100μA at 1A Full Scale |
Measuring Accuracy | < 1% of Full Scale in 10nA to 1μA ranges < 0,25% of Full Scale in 10μA to 1A ranges |
Reference Electrode | |
Input Impedance | > 1TΩ |
Input Capacitance | < 20pF (1m cable) |
Biasing Current | < 10pA at 25°C |
Common Mode Rejection | > 60dB full frequency response |
Voltage Range | ± 10V max or ± 50V max |
Input BNC | Grounded outer contact |
Polarization Capability | |
Voltage | ± 10V max |
Current | ± 1A max |
Voltage Resolution | 0,1mV |
Current Resolution | 10pA |
Accuracy | ± 0,2% & 0,1% (conversion at Full Scale) |
IR compensation | |
Positive Feedback Range | 2Ω to 100MΩ (depending on current range) |
Response Time | |
Potensiostatic Rise Time | < 1μs resistive load (1000Ω) |
Galvanostatic Rise Time | < 17μs resistive load (1000Ω) |
Meters and Interfaces | |
A/D Converter | 16 BIT |
D/A Converter | 16 BIT |
Temperature Meter | 0 to +100°C with PT1000 probe (0,1°C resolution and ± 0,2°C accuracy) |
Sampling Interval | 200µs |
Digital Interface | |
Connection | USB with full instrument control (baud rate 57600 – N – 8 – 1) |
Memory | EEPROM 64kB – SRAM 32kB |
Port Output | 8 external accessories |
I/O Port | 8 optional |
Cell Connections | |
Cables | 2, 3, 4 cables. BNC connector for Reference (Hi & Low) and PL258 for Working and Counter electrodes |
Power Supply and Dimensions | |
Voltage Mains | 115 or 230V AC ±10% 50/60Hz |
Power Consumption | 80VA max |
Dimensions (L x W x H) | 400 x 440 x 145mm |
Weight | 15kg |
Specifications are subject to change without previous notice
191/GPC | Grid power cable |
191/USB | USB cable A – B |
191/4BN4 | Set of 4 WE, RE and CE cables for tabletop potentiostats |
191/C4 | Set of 4 crocodile clips for WE, RE and CE 4mm banana plug |