| Operating mode |
Potentiostatic - Galvanostatic |
| Cell connections |
2,3,4 electrodes with guard |
| Counter electrode |
| Voltage output |
2 ranges : +/- 45 +/-16 Volt (user selectable)
Positive only output (mono position) for full current span on batteries
|
| Current output |
+/- 1.5 A max (45 V range)
+/- 4 A max (16 V range)
|
| Slew rate |
> 10 V / uS |
| Protection |
full thermal, overload and short-circuit protected |
| Working electrode |
| Current measure |
5A-5nA Full Scale in 10 ranges |
| Current resolution |
From 200 uA at 5 A F.S. to 200 fA at 5 nA F.S. |
| Measuring accuracy |
0.2 % +/- 0.1% f.s. |
| Reference electrodes |
|
| Input impedance |
> 1 TOhm |
| Input capacitanc |
< 20 pF (with 1 m cable and guard enabled) |
| Biasing current |
< 10 pA @ 25 C° |
| Common mode rejection |
> 50 dB full frequency response |
| Common mode voltage range |
+/- 45 V |
| Input BNC |
Outer contact grounded or guarded |
| Polarisation capabilities |
| voltage |
+/- 45 or +/-5 V |
| current |
+/-1.5 A at +/- 45 V compliance
+/-4.0 A at +/- 16 V compliance
|
| voltage resolution |
2 mV - 200 uV |
| max current resolution |
200 fA |
| accuracy |
+/- (0.2% +0.1% F.S.) worst case |
| Electrochemical tecniques |
| polarisation sweep |
Ramp, staircase, triangle, square wave |
| real part managing |
IR compensation or current interruption technique |
| limits of IR compensation |
0-400 % of the Rs |
| current interruption capabilities |
100 uS-5 mS time 1-255 on/off ratio |
| Frequency Response |
| Potentiostatic bandwidth (max) |
1 MHz (-3dB) resistive load (100 Ohm) |
| Galvanostatic bandwidth (max) |
100 KHz (-3dB) resistive load (100 Ohm) |
| Frequency bands |
1 MHz (Potentiostatic only) , 100 KHz, 20 KHz , 5 KHz |
| Meters and interfaces |
| Voltmeters & Ampere meters |
+/- 25000 counts 20% overload allowed |
| Temperature meter |
-200 + 500 C° with PT1000 probes
0.1C° resolution 0.2C° accuracy
|
| Sampling rate |
100 readout / sec (included data transmission) |
| Sampling techniques |
Synchronous sampling on current interrupt and on pulse voltammetry |
| Digital interface |
USB and RS 232 with full instrument control |
| Power supply and size |
| Voltage mains |
115 / 230 V ranges (+/- 10%) |
| Power consumption |
90 VA max |
| Size & weight |
400 x 430 x 135 mm (L x W x H) Kg 12
(rackmount brackets excluded)
|